Making Basic Measurements Using Spectrum Analyzer Mode

Tracking Unstable Signals Using Marker Track and the Maximum Hold and Minimum Hold Functions

Example: Use the marker track function to keep a drifting signal at the center of the display and monitor its change.

This example requires a modulated signal. An acceptable signal can be easily found by connecting an antenna to the spectrum analyzer input and tuning to the FM broadcast band (88 to 108 MHz). Set the spectrum analyzer center frequency for 100 MHz with a span of 20 MHz, an attenuator setting of 0 dB, and reference level setting of approximately 9 dBmV. Your circumstances may be slightly different, depending on building shielding and proximity to transmitters.

1.Connect an antenna to the spectrum analyzer input.

2.Press [PRESET], (FREQUENCY], 100 (MHz), ISPAN), 20 m.

N O T E

Use a different signal frequency if no signal is available at 100 MHz in your area.

3.Press @KiiiGKj, 9 (-1, ATTEN AUTO NAN , 0 @GZJ

4.Press m), SPAN ZOOM , 500 IkHz).

Notice that the signal has been held in the center of the display.

N O T E

If the signal you selected drifts too quickly for the spectrum analyzer to keep up with, use a wider span.