Spectrum Analyzer Mode Key Descriptions

Spectrum Analyzer Mode Functions

CHANNEL SPACING

C l e a r

allows the user to enter the channel spacing to set up the analyzer when using the measurement functions under the Power Menu softkey. When the power measurements are first accessed the initial value for channel spacing is

25 kHz. Once the value is changed, the new value will be saved through an instrument preset or power on. An error message will occur for invalid values of channel spacing and channel bandwidth.

If one of the power measurements is active and the channel spacing is changed, with PARAM AUTO selected, then

the coupled analyzer settings are immediately updated. If CONT MEAS is also selected, then another sweep is taken and the measured results are updated.

Front-Panel Key Access (MEAS/USERj

clears the current screen title or orefix. This softkey is accessed under the YZ,% Spc Clear softkey when using Change Title or Change Prefix.

Front-Panel Key Access (CAL), [CONFIGI), (jjl, (ml, or (SAVE)

CLEAR PARAM

clears all of the pulse parameters in the gate utility. It sets the value of pulse width, pulse repetition interval, and reference edge to zero and turns off the coupling of the pulse parameters to the resolution bandwidth, video bandwidth, and sweep time.

Front-Panel Key Access [SWEEP]

6-39