Preparing For Use

Turning on the Analyzer for the First Time

7.When the frequency and amplitude self-calibration routines have been completed successfully, store the correction factors by pressing

CAL STORE.

The self-calibration routines calibrate the cable TV analyzer by generating correction factors. The softkey CAL STOKE stores the correction factors in the area of cable TV analyzer memory that is stored when the cable TV analyzer is turned off; the cable TV analyzer will automatically apply these factors in future measurements. If CAL STOKE is not pressed, the correction factors remain in effect until the cable TV analyzer is turned off.

Performing the Tracking-Generator Self-Calibration Routine

For cable TV analyzers with Option 011, the tracking-generator self-calibration routine should be performed prior to using the tracking generator.

N O T E

Since the tracking generator calibration routine depends on the accuracy of the absolute amplitude level of the cable TV analyzer, the cable TV analyzer amplitude calibration should be done prior to using

CAL TRK GEN

l-15