Making Measurements Using Spectrum Analyzer Mode
Stimulus-Response Measurements
4.Put the sweep time of the analyzer into stimulus-response auto-coupled mode by pressing More I of 2 , then SWP CPLG SR SA until SR (stimulus-response mode) is underlined. Auto-coupled sweep times are usually much faster for swept-response measurements than they are for spectrum analyzer measurements.
N O T E
In the stimulus-response mode, the II (reactance versus resistance) of the OUT can determine the fastest rate at which the spectrum analyzer can be swept. To determine whether the analyzer is sweeping too fast, slow the sweep time and note whether there is a frequency or amplitude shift of the trace. Continue to slow the sweep time until there is no longer a frequency or amplitude shift.
I
4
REF 48.8 <lBmVAT 10 dB
PEAK / / 1
1 SPAN
I500 0 MHz
u I
\
FULL SPQN
ZERO SPAN
LAST SPAN
:ENTER 350 .O | MHz | | SPAN | 500.0 | MHz |
#RES BW | 3.0 MHz | VBW 1 MHz | SWP | 20.0 | msec |
Figure 5-6. Spectrum Analyzer Settings According to the Measurement Requirement