Spectrum Analyzer Mode Key Descriptions

Spectrum Analyzer Mode Functions

swaw

OPTIONS

SIGNAL

displays the number and description of the options installed in your analyzer, the instrument model number of the analyzer, the last five digits of the analyzer serial number, and the firmware revision. Pressing SHOW OPTIONS

changes the softkey label to EXIT SHOW Press EXIT SHOW to erase the SHOW OPTIONS function information.

Pressing SHOW OPTIONS displays the individual option numbers.

Front-Panel Key Access [WI

ID

activates an FFT marker that must be put on a signal to

 

verify that it is not being displayed at the wrong frequency

 

due to aliasing. Once the marker is on the signal press

 

SIGNAL ID again to initiate the signal identification

 

function. The signal should move half of a division to the

 

right for 2 seconds. A marker will be put at the frequency

 

that the signal should move to. If the marker appears on the

 

signal when it is shifted, then the frequency readout of the

 

signal is correct. Sweep time limitations may alter or stop

 

the function from executing.

 

If the marker or the signal is less than half of a division from

 

the right side of the display before SIGNAL ID is pressed,

 

then the routine may not function correctly and a warning

 

message can be displayed.

 