M a k i n g M e a s u r e m e n t s U s i n g S p e c t r u m A n a l y z e r M o d e
Using the
4.00 V/dIV
offset:2.062 v
,.000:1dc
| Figure 5.24. Placing the Gate Output During the Second Signal |
|
|
Item | Description of Items in Figure |
1Output from pulse generator 1.
2Output from pulse generator 2.
3Pulsed RF signal input to the spectrum analyzer.
4Gate output from
10.Set the resolution bandwidth to a value that is greater than 2 divided by the gate delay (110 ps) minus the signal delay (85 ps). The resolution bandwidth should be set to greater than 2 divided by 25 ps, or greater than 80 kHz. Press Isw_l, RES 3W , 100 m.