Spectrum Analyzer Mode Key Descriptions

Spectrum Analyzer Mode Functions

CLEAR WRITE A

erases any data previously stored in trace A and continuously displays any signals during the sweep of the analvzer. This function is activated at power on and by

Changing the trace mode of trace C to clear write or minimum hold can change the trace mode of trace A. If trace A is in clear-write mode or maximum-hold mode when trace C is changed to clear write or minimum hold, the trace mode of trace A is changed to store blank. The following table shows the trace mode of trace A before and after changing trace C to clear-write or minimum-hold trace mode.

)ITrace Mode of Trace A Trace Mode of Trace A

Changing the trace mode of trace A to clear write or maximum hold can change the trace mode of trace C. If trace C is in clear-write mode when trace A is changed to clear write or minimum hold, the trace mode of trace C is changed to minimum hold.

Front-Panel Key Access [TRACE)

6-40