Making Measurements Using Spectrum Analyzer Mode
Tracking Generator Unleveled Condition
When using the tracking generator, the message TG UNLVL may appear. The TG UNLVL message indicates that the tracking generator source power (SRC PWR ON OFF ) could not be maintained at the
The unleveled condition may be caused by any of the following:
•Start frequency is too low or the stop frequency is too high. The unleveled condition is likely to occur if the true frequency range exceeds the tracking generator frequency specification (especially the low frequency speciEcation). The true frequency range being swept may be significantly different than the start or stop frequency annotations indicate, depending on other
•Tracking peak may be required (use TRACKING PEAK ).
•Source attenuation may be set incorrectly (select SRC ATN MAN AUTO (AUTO) for optimum setting).
•The source power may be set too high or too low, use SRC PWR ON OFF to reset it.
•The source power sweep may be set too high, resulting in an unleveled condition at the end of the sweep. Use PWR SWP ON OFF to decrease the amplitude.