Spectrum Analyzer Mode Key Descriptions

Spectrum Analyzer Mode Functions

CARD + DLP

CARD

+ STATE

recalls into the analyzer memory a downloadable program (DLP) saved on the memory card. Before recalling data that was saved under a prefix other than the current prefix, change the current prefix to the prefix used when the data was saved. Pressing LOAD FILE is an alternate way to load program data from the memory card into analyzer memory. See “Saving and Recalling Data from the Memory Card” in Chapter 3 for more information. See also the Change Prefix softkey description.

Front-Panel Key Access [j]

recalls into analyzer memory a state saved on the memory card. CARD -+ STATE also displays the time and date when the state data was stored. Before recalling a state that was saved under a prefix other than the current prefix, change the current prefix to the prefix used when the state was saved.

If the windows display mode is being used, the instrument state can only be recalled into the active window.

Pressing LOAD FILE is an alternate way to load state data from the memory card into analyzer memory. See “Saving and Recalling Data from the Memory Card” in Chapter 3 for more information.

Front-Panel Key Access [RECALL)

6-24