Using Spectrum Analyzer Features

Using Limit-Line Functions

With the lower limit-line table format, one amplitude component (representing a lower limit-line segment) is specified per frequency/time component. The amplitude value is entered by pressing SELECT AMPLITUD , entering an amplitude value, and pressing a units key.

With the upper/lower limit-line table format, two amplitude components (one each for the upper and lower limit-line segments) can be specified per frequency or time component. It is not necessary to specify both an upper and a lower amplitude value. Specifying only upper amplitude values results in an upper limit line, but not a lower limit line. Omitting an amplitude point on one limit line does not affect the other limit line. The amplitude of the upper limit line is entered by pressing SELECT UPR AMPL , entering an amplitude value, and pressing a

units key. The amplitude of the lower limit line is entered by pressing SELECT LWR AMPL , entering an amplitude value, and pressing a units key.

With the mid/delta limit-line table format, two amplitude components

(one representing a mid-amplitude value, one representing a deviation [positive and negative values] from either side of this value) is specified per frequency component. If no deviation is entered, the deviation defaults to zero. The middle amplitude value is entered by pressing SELECT MID AMPL , entering an amplitude value, and pressing a units

key. The deviation is entered by pressing SELECT DLT AMPL , entering an amplitude value, and pressing a units key.

N O T E

Frequency or amplitude values that are not within the limit-line range will be modified. For example, a frequency value of 3 GHz will be modified to 1.9 GHz.

3 - 4 5