R

Chapter 12: Constraining the Core

In addition, the example design provides pad locking on the TBI for several families. This is included as a guideline only, and there are no specific I/O location constraints for this core.

TBI Input Setup/Hold Timing

Input TBI Timing Specification

PMA_RX_CLK0

PMA_RX_CLK1

rx_code_group[9:0]

tSETUP

tHOLD

Figure 12-2:Input TBI timing

tSETUP tHOLD

Figure 12-2and Table 12-1illustrate the setup and hold time window for the input TBI signals. These specify the worst-case data valid window presented to the FPGA device pins. There is only a 2 ns data valid window of guaranteed data presented across the TBI input bus. This must be correctly sampled by the FPGA devices.

Table 12-1:Input TBI Timing

Symbol

Min

Max

Units

 

 

 

 

tSETUP

2.00

-

ns

tHOLD

0.00

-

ns

Virtex-II, and Virtex-II Pro Devices

Figure 6-2illustrates the TBI input logic provided by the example design for the Virtex-II and Virtex-II Pro family. Although not illustrated, these families have input delay elements (always of a fixed delay). These are also automatically inserted by the Xilinx tools and are set to provide a zero-hold time. These input delays automatically meet input setup and hold timing on the TBI without any specific constraints.

Spartan-3, Spartan-3E, and Spartan-3A Devices

Figure 6-3, page 72 illustrates the TBI input logic provided by the example design for the Spartan-3 class family. DCMs are used on the pma_rx_clk0 and pma_rx_clk1 clock paths as illustrated. Phase-shifting is then applied to the DCMs to align the resultant clocks so that they correctly sample the 2 ns. TBI data valid window at the input DDR flip-flops.

The fixed phase shift is applied to the DCMs using the following UCF syntax.

INST "core_wrapper/tbi_rx_clk0_dcm" CLKOUT_PHASE_SHIFT = FIXED;

INST "core_wrapper/tbi_rx_clk0_dcm" PHASE_SHIFT = -10;

INST "core_wrapper/tbi_rx_clk0_dcm" DESKEW_ADJUST = 0;

170

www.xilinx.com

Ethernet 1000BASE-X PCS/PMA or SGMII v9.1

 

 

UG155 March 24, 2008

Page 170
Image 170
Xilinx 1000BASE-X manual TBI Input Setup/Hold Timing, 1Input TBI Timing Symbol Min Max Units