Data Types

Relay Function Codes (P111 Only)

Please note that the MultiRanger 200 offers more function codes.

If a device parameter returns a relay function code, that message is converted to a number and is then provided in the register. The numbers are shown in the following table:

MultiRanger 100

Control

Relay Function Code

Number

P111

General

OFF, relay not used

0

0

 

Undesignated Level Alarm

1

1

 

Low-Low Level Alarm

2

1 – LL

 

Low Level Alarm

3

1 – L

 

High Level Alarm

4

1 – H

 

High-High Level Alarm

5

1 – HH

 

Loss of Echo (LOE) Alarm

20

6

 

Transducer Cable Fault Alarm

16

7

Pump

Fixed Duty Assist

25

50

 

Alternate Duty Assist

30

52

Control

Communication

66

65

MultiRanger 200

 

 

 

 

 

 

Control

Relay Function Code

Number

P111

General

OFF, relay not used

0

0

 

Undesignated Level Alarm

1

1

 

Low-Low Level Alarm

2

1 – LL

 

Low Level Alarm

3

1 – L

 

High Level Alarm

4

1 – H

 

High-High Level Alarm

5

1 – HH

 

In Bounds Alarm

6

2

 

Out of Bounds Alarm

9

3

 

Rate of Level Change Alarm

12

4

 

Temperature Alarm

15

5

 

Loss of Echo (LOE) Alarm

20

6

 

Transducer Cable Fault Alarm

16

7

Pump

Totalizer

22

40

 

Flow Sampler

23

41

 

Fixed Duty Assist

25

50

 

Fixed Duty Backup

26

51

 

Alternate Duty Assist

30

52

Pump

Alternate Duty Backup

31

53

(con’t)

Service Ratio Duty Assist

35

54

 

Service Ratio Duty Backup

36

55

 

First In First Out (FIFO)

40

56

Control

Flush Valve

65

64

 

Communication

66

65

See P111 on page 134 of theParameter Reference section.