Operation

Setting the Default Auxiliary Reading

To maintain a constant variable display in the auxiliary reading area, set the default.

Example:

To leave the level reading on the screen and view the echo confidence in the auxiliary reading field, set the following parameter:

Parameter

 

Index

Value

 

Description

P730

G

805

 

Auxiliary field defaults to P805

Setting a Specific Auxiliary Reading

 

To display a second auxiliary reading, press

 

in RUN mode.

P

Example: To set

 

to display current temperature, go to P731:

P

Parameter

Index

Value

Description

P731

G

912

Shows P912–Transducer Temperature

Multiple Readings [MR 200 only]

During differential or average operation (P001 = 4 /5), the display scrolls sequentially through Point Numbers 1, 2, and 3. Point 3 is the difference between (or average of) Points 1 and 2.

Changing Number Scrolling Speed

Parameter

Index

Value

Description

P732

G

5

Hold each value for 5 seconds

See Parameter Indexing on page 37. All the instructions in the following procedures apply to the hand programmer and assume that the MultiRanger is activated.

Page 26

MultiRanger 100/200 – INSTRUCTION MANUAL

7ML19981FB06

Page 36
Image 36
Siemens Multiple Readings MR 200 only, Setting the Default Auxiliary Reading, Parameter Index Value Description, 912