P711 Echo Lock

Use this feature to select the measurement verification process.

Primary Index

 

Single Point Model

Dual Point Model

 

 

 

 

Global

Transducer

 

 

 

 

 

 

 

0

 

Off

 

 

 

 

 

 

Values

1

 

Maximum verification

 

 

 

 

 

2

*

Material agitator

 

 

 

 

 

 

 

 

 

3

 

Total lock

 

 

 

 

 

 

 

• P700 Max Fill Rate

 

 

• P701 Max Empty Rate

 

Related

P712 Echo Lock Sampling

 

 

• P713 Echo Lock Window

 

 

P820 Algorithm

 

 

 

 

 

 

If a material agitator (mixer) is used in the vessel monitored, set Echo Lock for maximum verification or material agitator to avoid agitator blade detection. Ensure the agitator is always ON while the MultiRanger is monitoring the vessel to avoid stationary blade detection.

When set for max verification or material agitator, a new measurement outside of the Echo Lock Window (P713) must meet the sampling criterion (P712).

For total lock, Echo Lock Window (P713) is preset to zero 0. The MultiRanger continuously searches for the best echo according to the algorithm chosen (P820). If the selected echo is within the window, the window is then centered about the echo. If not, the window widens with each successive shot until the selected echo is within the window. The window then returns to its normal width.

When Echo Lock is OFF , the MultiRanger responds immediately to a new measurement as restricted by the Max Fill / Empty Rate (P700 / P701); however, measurement reliability is affected.