P703 Emptying Indicator

The empty rate required to activate the LCD Emptying indicator ( ).

Primary Index

 

Single Point Model

Dual Point Model

 

 

 

Global

Transducer

 

 

 

 

Values

Range: 0.000 to 99.00 m (or equivalent depending on units)

 

 

Altered by

• P003 Maximum Process Speed

 

 

 

 

 

P005 Units

 

Related

P007 Span

 

 

• P701 Max Empty Rate

 

 

 

 

 

This value (in Units (P005) or % of Span (P007) per minute) is automatically set to 1/10 of the Max Empty Rate (P701).

P704 Rate Filter

Damps Rate Value (P707) fluctuations.

Primary Index

 

Single Point Model

 

Dual Point Model

 

 

 

 

 

Global

 

Transducer

 

 

 

 

 

 

 

 

 

0

 

Rate display not required

 

 

 

 

 

 

 

 

Filtered Output

 

 

 

 

 

 

 

1

 

Continuously filtered and updated

 

 

 

 

 

 

Values

Interval Output

 

 

 

 

 

 

2

 

1 minute or 50 mm (2 in)

 

 

 

 

 

 

 

 

 

3

 

5 minutes or 100 mm (3.9 in)

 

 

 

 

 

4

 

10 minutes or 300 mm (11.8 in)

 

 

 

 

 

5

 

10 minutes or 1000 mm (39.4 in)

 

 

 

 

 

 

Alters

P707 Rate Value

 

 

 

 

Altered by

P003 Maximum Process Speed

 

 

 

Related

P705 Rate Update Time/ P706 Rate Update Distance

 

 

 

 

 

 

Enter the time or distance interval over which the Rate Value is to be calculated before the display updates.

This is automatically altered along with Maximum Process Speed (P003).

This value automatically alters the Rate Update Time (P705) and / or Rate Update Distance (P706). Alternatively, these parameter values may be altered independently.

7ML19981FB06

MultiRanger 100/200 – INSTRUCTION MANUAL

Page 185