Auto Record ON and OFF Setpoints (P334 to P337)

Use Auto Record ON Setpoint (P334) and Auto Record OFF Setpoint (P335) to define the boundaries within which the level must be, for the resultant Echo Profile to be considered for an Auto Profile Record.

If ----is displayed for either P334 or P335, Auto Profile Records are saved regardless of

current level (subject to all other restrictions).

Enter the level value in Units (P005) or percent of Span (P007) as referenced to Empty (P006).

P334 Auto Record ON Setpoint

Enter the critical level which, in conjunction with Auto Record OFF Setpoint, defines the boundaries for Auto Profile Records to be saved.

Primary Index

Global

 

 

 

Values

Range: -999 to 9999

 

 

 

 

 

P335

Auto Record OFF Setpoint

Related

P336

Auto Record Filling / Emptying

 

P337 Auto Record LOE Time

 

 

 

 

P335 Auto Record OFF Setpoint

Enter the critical level which, in conjunction with Auto Record ON Setpoint, defines the boundaries for Auto Profile Records to be saved.

Primary Index

Global

 

 

 

Values

Range: -999 to 9999

 

 

 

 

 

P334

Auto Record ON Setpoint

Related

P336

Auto Record Filling / Emptying

 

P337 Auto Record LOE Time

 

 

 

 

Parameters

Page 162

MultiRanger 100/200 – INSTRUCTION MANUAL

7ML19981FB06