Soft-

 

 

 

ware

Date

 

Changes

Rev.

 

 

 

1.08

Nov. 11, 2004

• Improvements made as a result of Siemens

 

systems testing:

 

 

 

 

 

• SIMATIC PDM related: TVT display

 

 

 

improvement to prevent it from going

 

 

 

negative. MLFB number includes SMPI as

 

 

 

well as Siemens numbering. Change of some

 

 

 

enumerations in the Device Description.

 

 

• Device related: mA output no longer resets

 

 

 

when PAR_MA_FUNCTION (P201) is written

 

 

 

to. mA trims (P214, P215) now have the

 

 

 

property no Reset.

 

 

Enhancement: Device Identification

 

 

 

command allows Tag, Description, Message,

 

 

 

and Date.

 

 

 

1.09

Nov. 28, 2005

• Correction for manual Shaper TVT shift due

 

to change in velocity.

 

 

 

 

 

• Resolved issue of reading some floats as

 

 

 

decimal shifted integers.

 

 

• The submergence algorithm now

 

 

 

implemented with the thresholds defined.

 

 

• Changes to support single DUART.

 

 

 

 

1.10not released

1.11

Apr. 20, 2006

• Software modified to support replacement

component.

 

 

 

 

• Supports SIMATIC PDM version 6.02.

 

 

• Added support for true First echo processing

 

 

algorithm

 

 

• Support for 3rd Point Volume has been added

 

 

on P001 = DPA or DPD.

 

 

 

1.12

Aug. 15, 2008

• Many parameter ranges improved to match

product function and SIMATIC PDM.

 

 

 

 

• P110 = OFF removed from SIMATIC PDM

 

 

• Improved text labels for P770

 

 

 

Appendix H: Software Rev.

7ML19981FB06

MultiRanger 100/200 – INSTRUCTION MANUAL

Page 253

Page 263
Image 263
Siemens 200 Systems testing, Simatic PDM related TVT display, Improvement to prevent it from going, Property no Reset