Installation

Installation

Notes:

Installation must only be performed by qualified personnel, and in accordance with local governing regulations.

This product is susceptible to electrostatic shock. Follow proper grounding procedures.

All field wiring must have insulation suitable for at least 250 V.

Hazardous voltage present on transducer terminals during operation.

DC terminals shall be supplied from an SELV source in accordance with IEC 1010-1 Annex H.

The non-metallic enclosure does not provide grounding between conduit connections. Use grounding type bushings and jumpers.

Mounting

Mounting Locations

Recommended

Ambient temperature is always within -20 to 50 °C (-5 to 122 °F)

MultiRanger display window is at shoulder level, unless most interaction is through a SCADA system

Easy access for hand programmer is provided

Cable length requirements are minimal

Mounting surface is free from vibration

Leave is sufficient room to swing unit lid open and have clear access.

A place for a laptop computer is provided for on-site Dolphin Plus configuration

Avoid

Exposure to direct sunlight. (Provide a sun shield to avoid direct sunlight.)

Proximity to high voltage/current runs, contacts, SCR or variable frequency motor speed controllers

Page 8

MultiRanger 100/200– INSTRUCTION MANUAL

7ML19981FB06