Channel Monitoring

MR 200

Open

 

 

 

H Flume

Transducer

Flow

 

 

 

Point of Measurement

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Plan View

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

P006

 

 

 

D

 

 

 

 

 

 

P007

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Front View

 

Side View

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter

Index

 

 

Value

 

 

 

 

 

 

P600

G

3–H Flume

 

 

 

 

 

 

P602

1

 

Flume height (D)

 

 

 

 

 

 

P603

G

Maximum Head (preset = P007)

P604

G

Maximum Flow

 

 

 

 

 

 

P606

G

Time Units

 

 

 

 

 

 

Sized by maximum depth of flume

Approach is preferably rectangular, matching width and depth for distance 3 to 5 times the depth of the flume

May be installed in channels under partial submergence (ratio of downstream level to head). Typical errors are:

1% @ 30% submergence

3% @ 50% submergence

For rated flows under free flow conditions, the head is measured at a point downstream from the flume entrance

Flume Size

(Diameter in feet)

0.5

0.75

1.0

1.5

2.0

2.5

3.0

4.5

Point of Measurement

cminches

5

7

9

14

18

23

9

28

10¾

41

16¼

H flumes come with a flat or sloping floor. The same flow table can be used because error is less than 1%.