With P112 and P113, you can set a high alarm (P112 > P113) or a low alarm (P112 < P113).

This shows a high alarm:

Parameter

P111

P112

P113

P110

Index

5

5

5

5

Value Description

5Alarm on temperature

45 ON setpoint at 45 °C

43 OFF setpoint at 43 °C

1 Take the temperature reading from Transducer One

The temperature source can be the temperature sensor built into the transducer or an external TS-3, as set by P660.

Loss of Echo (LOE)

Parameter

Index

Value

Description

P110

5

1

Alarm on LOE for Transducer One

P111

5

6

Alarm on LOE

P070

G

0.5

Trip alarm when 0.5 minutes (30 seconds)

 

 

 

pass without detecting a valid echo.