Calibrates Empty (P006) if the reported level is consistently high or low by a fixed amount.

Primary Index

 

Single Point Model

 

Dual Point Model

 

 

 

 

 

Global

Transducer

 

 

 

 

 

 

Values

Range: -999 to 9999

 

 

 

 

 

 

 

 

 

 

P006 Empty

P652

Offset Correction

Related

P062 Offset Reading

P664

Temperature

 

P605 Zero Head

 

 

 

 

 

 

 

 

 

Before using this feature, verify the following parameters are correct:

Empty (P006)

Temperature (P664)

Offset Reading (P062)

Zero Head Offset (P605), if using OCM

Offset Calibration

Begin with a steady level.

1.Press TRANSDUCER to display the calculated reading.

2.Repeat Step One at least five times to verify repeatability.

3.Measure the actual reading (use tape measure).

4.Enter the actual value.

The deviation between the entered Empty (P006) value and the calibrated Empty value is stored in Offset Correction (P652).

P651 Sound Velocity Calibration

Changes the speed of sound constant.

Primary Index

Single Point Model

Dual Point Model

 

 

 

Global

 

Transducer

 

 

 

 

 

Values

Range: -999 to 9999

 

 

 

 

 

Related

• P653

Velocity

 

• P654

Velocity at 20°C

 

 

 

 

 

 

 

Condition for use of this feature

The acoustic beam atmosphere is other than air

The acoustic beam atmosphere temperature is unknown

The Reading accuracy is acceptable at higher material levels only

For best results, calibrate with the level at a known value near empty.