Parameters

P902 Watchdog

Press ENTER to put the CPU into an infinite loop to test the watchdog timer.

On successful completion (10 seconds) the RUN mode is entered and the MultiRanger is reset. Programming is kept and the unit responds as if there had been a power failure.

P903 Display

Press ENTER to activate the display test.

All LCD segments and symbols are temporarily displayed.

P904 Keypad

Press ENTER , then press each keypad key in the following sequence:

P P

As each key is pressed, the associated keypad number is displayed. On successful test completion, PASS is displayed. FAIL is displayed if a key is pressed out of sequence or the programmer keypad malfunctions.

P905 Transmit Pulse

Press ENTER to supply repeated transmit pulses, at the frequency entered, to the transducer and / or view the transducer operating frequency (automatically altered by P004 Transducer) for the Point Number displayed.

Primary Index

Single Point Model

Dual Point Model

 

 

Global

Transducer

 

 

 

 

Values

Range: 42kH to 46KH (view only)

 

 

 

 

Altered By

• P004 Transducer

 

 

 

 

P906 RS-232 Port

Press ENTER to test the RS-232 port on the RJ-11.

An external device must be connected to the RS-232 port for this test. On successful completion, PASS is displayed, otherwise it is FAIL.