P927 Distance Measurement

The distance between the surface and the transducer face (displays only as % of Empty).

Primary Index

Transducer

 

 

Values

Range: 0.000 to 9999% (Displays as % of Empty) (view only)

 

 

 

Related

P005 Units

P006 Empty

 

 

 

 

Use P923 unless the distance information is required in percent.

Master Reset (P999)

This feature resets all parameters to original values.

Primary Index

Single Point Model

Dual Point Model

 

 

Global

Transducer

 

 

 

 

Values

See below.

 

 

 

 

Use this feature prior to initial programming if arbitrary Parameter Values were used during a bench test, or after upgrading the software. Following a Master Reset, complete reprogramming is required.

To perform a Master Reset, access P999 and press CLEAR . C.ALL displays until the reset is complete.

In dual point units, both points can be reset at the same time by setting the index to 00

and pressing the CLEAR keys.

CAUTION: be careful when using this feature. All data for all points will be reset. For convenience, be sure to record the values you want to re-enter.