Siemens 200 Test P900 to P913, P852 Short Shot Range, P900 Software Revision Number, P901 Memory

Models: 200

1 282
Download 282 pages 42.81 Kb
Page 225
Image 225

P852 Short Shot Range

Enter the maximum distance in Units (P005) to be measured using short shot echoes.

Primary Index

 

Single Point Model

Dual Point Model

 

 

 

 

Global

 

Transducer

 

 

 

 

 

Values

Range: 0 to 10 m, or equivalent depending on units

 

 

 

 

Altered By

P004 Transducer

 

 

 

 

 

 

P840 Short Shot Number

 

 

P842

Short Shot Frequency

 

Related

P844

Short Shot Width

 

 

P850

Short Shot Bias

 

 

P851

Short Shot Floor

 

 

 

 

 

 

This feature is automatically altered when Transducer (P004) is altered.

Test (P900 to P913)

Test Parameters are intended for use by Siemens Milltronics Service personnel.

P900 Software Revision Number

View the EPROM Rev. #.

Primary Index

Global

 

 

Values

Range: 00.00 to 99.99 (view only)

 

 

P901 Memory

Press ENTER to activate the MultiRanger memory test.

Primary Index

Global

 

 

 

 

 

Display: view only

 

 

 

 

 

PASS

 

(memory test successful)

Values

 

 

 

F1

 

RAM

 

 

 

 

 

F3

 

FLASH data

 

 

 

 

 

F4

 

FLASH code

 

 

 

 

7ML19981FB06

MultiRanger 100/200 – INSTRUCTION MANUAL

Page 215

Parameters

Page 225
Image 225
Siemens 200 instruction manual Test P900 to P913, P852 Short Shot Range, P900 Software Revision Number, P901 Memory