Siemens 200 P061 Convert Reading MR, P062 Offset Reading, P607 Flowrate Decimal, Range -999 to

Models: 200

1 282
Download 282 pages 42.81 Kb
Page 138
Image 138

Parameters

Alters

 

P607 Flowrate Decimal

Altered by

 

P051 Maximum volume

Related

 

P920 Reading Measurement

 

 

 

 

In RUN mode, the decimal position adjusts to prevent the number of digits from exceeding the display capabilities. To keep the decimal place from shifting, reduce the number of decimal places to that shown at 100%.

Example:

If 100% is 15m, use two decimal places for sample readings of 15.00 or 12.15.

P061 Convert Reading [MR 200]

Multiplies the current value by the specified amount to allow for scaling.

Primary Index

Level

Values

Range: -999 to 9999

 

 

Preset: 1.000

 

Related

P920 Reading Measurement

 

 

 

Examples:

If the measured value is in feet, enter 0.3333 to display the number of yards

For simple linear, volume conversions set POO5 to 1 (meters) and then enter the volume measurement per unit to get the correct conversion. For example, if the reservoir contains 100 litres per vertical meter, use 100 to get the reading in litres.

Notes:

This method does not calculate volume. It must not be used in place of the volume parameters if any volume dependent features (such as pump efficiency) are used. To calculate true volumes see Volume (P050 to P055).

Avoid entering a value that, when multiplied by the maximum current Reading, exceeds the display capabilities. If value exceeds four digits, EEEE is shown.

P062 Offset Reading

Adds the specified value to the level reading, usually to reference the reading to sea level or another datum level.

Primary Index

Level

Values

Range: -999 to 9999

 

 

Preset: 0.000

 

Related

P920 Reading Measurement

 

 

 

The operation of the device is not affected by the Offset Reading. This value is used for display purposes only. All control measurements are still referenced to Empty.

Page 128

MultiRanger 100/200 – INSTRUCTION MANUAL

7ML19981FB06