Page 212 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06
mmmmm
Paramete rs
Advanced Shot Adjustment (P840 to P852)
These parameters are for Siemens Milltronics service personnel only.

P840 Short Shot Number

The number of short shots to be fired (and results averaged) per transmit pulse.

P841 Long Shot Number

Enter the number of long shots to be fired (and results averaged) per transmit pulse.

This value is automatically altered by Maximum Process Speed (P003).
Primary Index
Single Point Model Dual Point Model
Global Transducer
Values
Range: 0 to 100
Preset: 1
Related
P841 Long Shot Number
P842 Short Shot Frequency
P844 Short Shot Width
P850 Short Shot Bias
P851 Short Shot Floor
P852 Short Shot Range
Primary Index
Single Point Model Dual Point Model
Global Transducer
Values
Range: 0 to 200
Preset: 5
Altered By P003 Maximum Process Speed
Related
P840 Short Shot Number
P843 Long Shot Frequency
P845 Long Shot Width