P735 Backlight

Controls the LCD backlighting.

Primary Index

Global

 

 

 

 

 

 

0

 

Off

Values

 

 

 

1

*

On

 

 

 

 

 

2

 

Keypad activated

 

 

 

 

The backlight can be forced on or off, or be controlled by a programmer, in which case it will turn OFF 30 seconds after the last key is pressed.

P737 Primary Reading [MR 200]

The reading shown on the primary reading display when in RUN mode.

Primary Index

Global

 

 

 

Range: 0 to 3

 

 

 

 

Values

1

*

Default reading (P920) based on operation (P001)

 

 

 

2

 

LCD totalizer (P322, P323)

 

 

 

 

 

 

 

3

 

Automatically toggle between 1 and 2

 

 

 

 

 

LCD Totalizer (P322 and P323)

Related

P732 Display Delay

 

P920 Reading Measurement

 

 

 

 

When this value indicates TOGGLE, then both readings (default and totalizer) are shown in the time specified in display delay (P732).

P741 Communications Timeout

The maximum time allowed between receiving a request and transmitting the response.

Primary Index

Port

Range: 0 to 60 000 milliseconds

Values

Preset: 5 000 ms

If the maximum time is exceeded, no response will be transmitted, and the action required may not be completed.

7ML19981FB06

MultiRanger 100/200 – INSTRUCTION MANUAL

Page 193

Parameters

Page 203
Image 203