Backup Level Override

Use this feature to override the ultrasonic reading by a discrete input such as a contacting point device. The ultrasonic reading will be fixed at the programmed switch level until the discrete input is released.

The ultrasonic device makes decisions based on the override values.

P064 Reading Override Enable

Sets the discrete input to act as the source for a level reading override.

Primary Index

Transducer

Values

0

*

OFF: No override.

 

 

 

1-2

 

ON: Number = discrete input of override signal

 

 

Related

P065

Reading Override Value

P270

Discrete Input Function

 

 

 

 

 

P065 Reading Override Value

This value is substituted for the current reading when the selected discrete input is enabled and activated.

Primary Index

Transducer

Values

Range: 0.0 to 99.00 m (or equivalent depending on units)

 

Preset: 0.00

Alters

Current reading

 

P001 Operation

 

P005 Units

Related

P006 Empty

 

P007 Span

 

P064 Override Enable

 

 

 