P838 Auto False-Echo Suppression Distance

Defines the range of Auto False-Echo Suppression (P837) to use for ignoring false echoes. (Units are defined in P005.)

Values

Range

Maximum range:

(depends on model)

0.000 to 15 m (50 ft)

 

Default

1.000 m (3.28 ft)

 

 

 

Determine the actual distance from the transducer face to the material surface. Subtract 0.5 m from this distance, and enter the result.

Set Up:

1.Perform this function when the vessel is empty or nearly empty.

2.Determine actual distance from transducer face to material level.

3.Select P838 and key in [distance to material level minus 0.5 m].

4.Press ENTER .

5.Select P837.

6.Press 2 and then press ENTER . P837 will revert to 1 (use Learned TVT) automatically after a few seconds.

Display after Auto False Echo Suppression

 

 

 

TVT curve

 

 

 

material

(learned)

 

 

level

 

 

 

 

 

 

(db)

 

Level

false

echo

 

 

 

Distance (meters)

P839 TVT Hover Level

Defines (in percent) how high the TVT curve is placed above the profile, relative to the largest echo. When MultiRanger 100/200 is located in the center of the vessel, lower this parameter to prevent multiple echo detections.

Values

Range

0 to 100%

 

 

Default

33 (%)

 

 

 

 