Page 234 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06
mmmmm
Appendix C: Troubleshooting
Measurement Difficulties
If the Failsafe Timer (P070) expires due to a measurement difficulty, LOE flashes
alternately with the last known Reading. In rare cases, the MultiRanger may lock on to a
false echo and report a fixed or wrong reading.

Flashing LOE Display

The loss of echo (LOE) display appears when the echo confidence is below the threshold
value set in P805 Echo Confidence.
LOE occurs when:
The echo is lost and no echo is shown above the ambient noise. See low echo
confidence (P805) and low echo strength (P806).
Two echoes are too similar to differentiate. See low echo confidence (P805) and low
echo strength (P806).
If LOE is displayed, check the following:
Surface monitored is within the transducer maximum range
Transducer model (P004) matches the transducer used
Transducer is located and aimed properly
Transducer is not submerged without a submergence shield

Adjust Transducer Aiming

See the transducer manual for range, mounting, and aiming details. For optimum
performance, adjust transducer aiming to provide the best Echo Confidence (P805) and
Echo Strength (P806) for all material levels within the measurement range.
The most efficient method of checking echoes is with Siemens Milltronics Dolphin Plus
software.
Displaying Echoes
Use Dolphin Plus to graphically display
the echo profile at the installation.
Interpret the echo profile and change
relevant parameters.
Editing Parameters
Edit the parameter values. Use F1
to get online help at any time.