Siemens 200 Checking Volume Characterization MR, Checking OCM Flow Characterization MR, Press

Models: 200

1 282
Download 282 pages 42.81 Kb
Page 96
Image 96

Testing the Configuration

3.Press the ARROW or to adjust the simulated rate of rise or fall, based on the chart on page 86. The maximum rate is 4% of Span / second.

The effect of the ARROW key is determined by the state (rate of rise or fall) immediately before the key is pressed.

Action

State (prior to pressing key)

Effect

 

 

 

 

Stop

Rise at 1% of Span / second

 

 

 

 

Rise at 1% of Span / second

Rise at 4% of Span / second (max.)

 

 

 

Press

Rise at 4% of Span / second (max.)

No effect

 

 

Fall at 1% of Span / second

Stop

 

 

 

 

 

Fall at 4% of Span / second

Fall at 1% of Span / second

 

 

 

 

Stop

Fall at 1% of Span / second

 

 

 

 

Rise at 1% of Span / second

Stop

Press

 

 

Rise at 4% of Span / second (max.)

Rise at 1% of Span / second

 

 

 

 

Fall at 1% of Span / second

Fall at 4% of Span / second (max.)

 

 

 

 

Fall at 4% of Span / second (max.)

No effect

 

 

 

When the level rises to 100% or falls to 0%, it reverses direction at the same rate.

Checking Volume Characterization [MR 200]

To confirm universal volume calculations (P050 = 9, 10) are correct:

1.Go to P920.

2.Key in a level associated with a known volume.

3.Press ENTER .

4.Check the returned volume against the manufacturer’s chart.

5.Change parameters P054 and P055, as required.

6.Repeat steps 2 to 5 until the volume curve is verified.

Checking OCM Flow Characterization [MR 200]

To confirm universal flow calculations (P600 = 4, 5) are accurate:

1.Go to P925.

2.Enter a level with a known flow.

3.Press ENTER .

4.Check the returned flow against the manufacturer’s chart.

Page 86