P821 Spike Filter

Dampens spikes in the echo profile to reduce false readings.

Primary Index

 

Single Point Model

Dual Point Model

 

 

 

 

 

Global

Transducer

 

 

 

 

 

 

 

Values

0

 

 

Off

 

 

 

 

 

 

1

 

*

On

 

 

 

 

 

 

 

 

 

 

 

 

P820 Algorithm

 

Related

• P822 Narrow Echo Filter

 

 

P823 Reform Echo

 

 

 

 

 

• P825 Echo Marker Trigger

 

 

 

 

 

 

 

Use P821 if interference spikes are on the long shot Echo Profile display.

P822 Narrow Echo Filter

Filters out echoes of a specific width.

Primary Index

 

Single Point Model

Dual Point Model

 

 

 

 

Global

 

Transducer

 

 

 

 

 

Values

0 = OFF (preset), nearest acceptable value returned

 

 

 

 

greater = wider

 

 

 

 

 

 

 

 

 

P820

Algorithm

 

Related

P821

Spike Filter

 

P823 Reform Echo

 

 

 

 

• P825

Echo Marker Trigger

 

 

 

 

 

 

Use this for transducer acoustic beam interference (e.g. ladder rungs). Enter the width of false echoes (in ms) to be removed from the long shot Echo Profile. When a value is keyed in, the nearest acceptable value is entered.

7ML19981FB06

MultiRanger 100/200 – INSTRUCTION MANUAL

Page 205

Parameters

Page 215
Image 215
Siemens 200 P821 Spike Filter, P822 Narrow Echo Filter, Dampens spikes in the echo profile to reduce false readings