Parameters

P705 Rate Update Time

The time period (in seconds) over which the material level rate of change is averaged before Rate Value update.

Primary Index

 

Single Point Model

Dual Point Model

 

 

 

Global

Transducer

 

 

 

 

Values

Range: 0.000 to 9999

 

 

 

 

 

Related

P707 Rate Value

 

 

 

 

 

P706 Rate Update Distance

The material level change (in metres) to initiate a Rate Value update.

Primary Index

Single Point Model

Dual Point Model

 

 

Global

Transducer

 

 

 

 

Values

Range: 0.000 to 9999

 

 

 

 

Related

• P707 Rate Value

 

 

 

 

P707 Rate Value

The rate of material level change (in Units (P005) or % of Span (P007) per minute).

Primary Index

 

Single Point Model

Dual Point Model

 

 

 

Global

Transducer

 

 

 

 

Values

Range: -999 to 9999 (view only)

 

 

 

 

 

Altered By

P704 Rate Filter

 

 

 

 

 

Related

P005 Units

 

P007 Span

 

 

 

 

 

 

 

A negative rate indicates the vessel is emptying.

This is the value displayed when is pressed in the RUN mode as described in the Readings in RUN Mode chart on page 23.