Siemens 200 instruction manual P203 mA Output Value / Transducer

Models: 200

1 282
Download 282 pages 42.81 Kb
Page 158
Image 158

Enter the Point Number the mA output is based on. This value depends on whether mA function (P201) is set as transducer or mA input.

For a single point device, if P201 uses a transducer, parameter P202 can only be altered if P001 (Operation) is set for DPD or DPA.

P202 can be set to a particular point or range of points. When set to a range of points, the mA output will be the average of the readings from all in service transducers in the range. Out of service transducers will be ignored.

If P202 mA Output Allocation End contains the value 0, then only 1 transducer has been assigned to the mA output.

P203 mA Output Value / Transducer

Displays current mA output value for the Point Number.

Primary Index

Level

 

 

Values

Range: 0.000 to 22.00 (view only)

 

 

This displays as an Auxiliary Reading when key is pressed in the RUN mode and does not include adjustments made using Trim features (P214 / P215).

P203 holds the mA value associated with a transducer. This value is used for run mode display of the mA output. When only one mA output is associated with a transducer, this parameter holds the value of that mA output. When more than one transducer is associated with a mA output (set via P202), P203 holds the mA output calculated from the resulting average of the associated transducer readings. When more than one mA output is associated with a transducer, P203 holds the value of the first mA output associated

with the transducer.

Note: This parameter is applicable only if any mA output has the transducer

Point Number as its input source (see P201 and P202).

Parameters

Page 148

MultiRanger 100/200 – INSTRUCTION MANUAL

7ML19981FB06

Page 158
Image 158
Siemens 200 instruction manual P203 mA Output Value / Transducer, Displays current mA output value for the Point Number