Paramete rs

P303 Temperature, Sensor Minimum

Shows the lowest temperature encountered, as measured by the TS-3 Temperature

Sensor (if applicable).

Press CLEAR keys to reset the log after an open circuit on the transducer wiring.
Record Readings (P304 and P305)
This identifies the occurrence of the record high and low level readings. Press CLEAR
keys to reset these values once the installation is working correctly.

P304 Reading Max

Shows the highest Reading calculated (in normal Reading units or %).

P305 Reading Min

Shows the lowest Reading calculated (in normal Reading units or %).

Pump Records (P309 to P312)
These features to identify pump usage and if the associated Relay Function (P111) is set
for any pump control feature. The value is that of the pump connected to the associated
terminals.
Enter a value to set the current record to that value. Use this if a pump is added with a
known number of hours logged, or the value can be reset to zero 0 after maintenance.
Primary Index Global
Values
Range: - 50 to 150°C (view only)
Preset: 150°C
Related P302 Temperature, Sensor Max
Primary Index Level
Values Range: -999 to 9999 (view only)
Related P305 Reading Min
Primary Index Level
Values Range: -999 to 9999 (view only)
Related P304 Reading Max