Parameters

P603 Maximum Head [MR 200]

The level value associated with Maximum Flow, in Units (P005).

Primary Index

 

Single Point Model

Dual Point Model

 

 

 

Global

Transducer

 

 

 

 

Values

Range: -999 to 9999

 

 

 

 

Preset: Span (P007) value

 

 

 

 

 

 

 

Altered By

P005 Units

 

• P600 Primary Measuring Device

 

 

 

 

 

Related

P604 Maximum Flow

 

P605 Zero Head

 

 

 

 

 

 

 

This represents the highest head level supported by the PMD and works in conjunction with Maximum Flow (P604) to define the highest point in the exponential curve. Use it when the Primary Measuring Device (PMD) requires a maximum head and flow reference point. This would include Exponential, Palmer Bowlus Flume, H-Flume, and Universal breakpoints.

P604 Maximum Flow [MR 200]

The maximum flowrate associated with Maximum Head (P603).

Primary Index

 

Single Point Model

Dual Point Model

 

 

 

Global

Transducer

 

 

 

 

Values

Range: -999 to 9999

 

 

 

 

Preset: 1000

 

 

 

 

 

 

Altered By

P600 Primary Measuring Device

 

 

 

 

 

P603 Maximum Head

 

Related

P606 Time Units

 

 

P925 Flow Measurement

 

 

 

 

 

This represents the flow at the highest head level supported by the PMD. and works in conjunction with Maximum Head (P603) to define the highest point in the exponential curve. Use it when the Primary Measuring Device (PMD) requires a maximum head and flow reference point. This would include Exponential, Palmer Bowlus Flume, H-Flume, and Universal breakpoints.

Also use this parameter with Time Units (P606) to define the flowrate units. The limitation of four digits is for the LCD only, and the flowrate value is available with greater precision through communications.