P820 Algorithm

Chooses the algorithm to generate the measured value from the profile.

Primary Index

 

Single Point Model

 

Dual Point Model

 

 

 

 

 

 

 

Global

 

 

 

Transducer

 

 

 

 

 

 

 

 

 

 

 

Values

1

 

 

 

ALF = flat Area, Largest, and First average

 

 

 

 

 

 

 

 

2

 

 

 

A = flat Area only

 

 

 

 

 

 

 

 

 

3

 

 

 

L = flat Largest only

 

 

 

 

 

 

 

 

 

4

 

 

 

F = flat First only

 

 

 

 

 

 

 

 

5

 

 

 

AL = flat Area and Largest average

 

 

 

 

 

 

 

6

 

 

 

AF = flat Area and First average

 

 

 

 

 

 

 

7

 

 

 

LF = flat Largest and First average

 

 

 

 

 

 

 

8

 

*

 

bLF = smooth Largest or First

 

 

 

 

 

 

 

9

 

 

 

bL = smooth Largest only

 

 

 

 

 

 

 

 

10

 

 

 

bF = smooth First only

 

 

 

 

 

 

 

 

 

12

 

 

 

tF = true First only

 

 

 

 

 

 

 

 

 

 

P805

Echo Confidence

 

 

P821

Spike Filter

 

Related

P822

Narrow Echo Filter

 

 

P823 Reform Echo

 

 

P825

Echo Marker Trigger

 

 

 

 

 

 

 

 

 

Use this to select the algorithm(s) the Sonic Intelligence® echo selection is based on. Use P805 Echo Confidence (page 202) to determine which algorithm gives the highest confidence under all level conditions. If the wrong echo is processed, observe the echo processing displays and select an alternate algorithm, either by entering the numeric value desired, or as below:

1.Press MEASURE to display the Auxiliary Function symbol.

2.Press ARROWS to access the desired Reading display symbols.

3.Press ENTER when the required algorithm is displayed.