P803 Shot / Pulse Mode

Determines what type of ultrasonic shots are fired.

Primary Index

 

Single Point Model

Dual Point Model

 

 

 

 

 

Global

Transducer

 

 

 

 

 

 

 

Values

1

 

 

Short

 

 

 

 

 

 

2

 

*

Short and long

 

 

 

 

 

 

 

 

 

 

 

 

P006 Empty

 

Related

 

P805 Echo Confidence

 

 

P804 Confidence Threshold

 

 

 

 

 

• P852 Short Shot Range

 

 

 

 

 

 

 

Increases MultiRanger response when the monitored surface is close to the transducer face. Select short and long to have short and long acoustic shots fired for each measurement, regardless of the transducer to surface distance. Select short to have only short shots fired if the Echo Confidence (P805) produced by a short shot exceeds the short Confidence Threshold (P804) and the monitored surface is always within the Short Shot Range (P852).

P804 Confidence Threshold

Determines which echoes are evaluated by software.

Primary Index

Single Point Model

Dual Point Model

 

 

Global

Transducer

 

 

 

 

Values

Range: 0 to 99:0 to 99

 

 

 

Preset: 10:5

 

 

 

 

 

 

Related

• P805 Echo Confidence

 

 

 

 

The short and long shot Confidence Thresholds are preset to 10 and 5 respectively. When Echo Confidence (P805) exceeds the Confidence Threshold, the echo is evaluated by Sonic Intelligence®. Values are entered as two numbers separated by a decimal point. The first number is the short shot confidence and the second number is the long shot confidence.

Note: The decimal point is replaced with a colon (:) on the display.