Operating manual
5.
CALIBRATION | a canned cycle |
MANUAL (JOG) MODE. TOOL | Automatic calibration with a probe and |
CNC 8070
(SOFT V03.0X)
Definition of data
To define the data, place the focus on the relevant data, key in the desired value and press [ENTER]. To change icons, place the focus on it and press [SPACE].
Data | Meaning |
|
|
T | Tool to be calibrated. |
|
|
D | Tool offset to be calibrated. |
|
|
Ds | Safety distance. |
|
|
F | Probing feedrate. |
| If not defined, the movements are carried out at the default feedrate, |
| set by the machine manufacturer. |
|
|
PRB1MAX | Probe position. |
··· | The values defined here are only taken into account during the |
PRB2MIN | calibration cycle; they do not modify the machine parameter values. |
|
|
Tool calibration steps
To calibrate the length, follow these steps.
1.Select the tool and the offset to be calibrated. It is not necessary to insert the tool into the spindle; the CNC carries out this operation if necessary.
2.Define the data defining the calibration.
3.Press the [CYCLE START] key to start the calibration. The CNC calibrates the tool making all the necessary movements; there is no need to manually approach the tool. If necessary, the CNC makes the tool change.
4.After the calibration It updates the tool table data.
90