Siemens SITRANS FUS1010 NEMA-4X/ IP 65 manual Datalogger Control Menu Structure

Models: SITRANS FUS1010 NEMA-4X/ IP 65

1 232
Download 232 pages 51.98 Kb
Page 153
Image 153

Section 4

1010NFM-3J

 

 

 

4.4THE DATALOGGER CONTROL MENU

The Datalogger Setup menu in the Channel Setup menu provides the Datalogger controls for the meter’s measurement channels. It allows you to enable usage, select data items/ alarm events, a logging interval and a destination for your Datalogger reports. While the Datalogger Setup menu is measurement channel specific, this Datalogger Control menu provides global control functions. This means that the settings made here apply to all measurement channels, meter types, operating modes, etc. This is possible because the meter stores logged data in a single file. This is significant for dual- channel systems, since the logged data from both channels are combined. Therefore, select the Site ID item for each channel to be logged.

Datalogger Control allows you to select a Circular Memory mode that will over-write the oldest Datalog- ger data automatically when the Datalogger memory becomes full due to the data compression scheme employed. Note that this is only available for dual-channel systems. The [Est Log Time Left] menu view-only menu cell shows an estimate of the hours and minutes of logging time remaining. This only applies to non-circular datalogging. Selecting Circular memory blanks this field. For convenience sake, the Display Datalogger command is essentially a duplicate of the menu cell in Datalogger Setup. It sends Datalogger data to the graphic screen with or without line wrapping. For dual-channel systems with logging enabled on both channels, a screen dump from menu will show data from both channels. The Output Datalogger command sends data to an external device via the RS-232 serial port. Data transmitted from this menu will be from both channels of a dual-channel system. The Clear Datalogger command erases the entire Datalogger file. For dual-channel systems, you are cautioned not to use this command if your intention is to clear the logged data of only one channel.

To access the Datalogger Control Menu press

DATALOGGER CONTROL MENU STRUCTURE

Datalogger ControlÖ

Display DataloggerÖ

Off

 

 

ÚLine Wrap

 

 

No Line Wrap

Output Datalogger Ö

ÚNo

 

 

Yes

Circular Memory

Ö

ÚNo

 

 

Yes

Est. Log Time Left Ö

Ú—:—

Clear Datalogger

Ö

ÚNo

 

 

Yes

4.4.1 DISPLAY DATALOGGER

This menu cell allows you to send the Datalogger contents to the display screen. This command is effective only after a successful install. You can set the report to scroll on the screen with or without line-wrap. Selecting line wrap, forces a line feed after approximately 40 characters. In addition, you have to enable datalogging and then select items in the Datalogger Setup menu. Note that this com- mand transmits the data from both channels.

To send Datalogger contents to the display screen:

To access the Display Datalogger option list press

Scroll cursor to either [Line Wrap] or [No Line Wrap] by pressing 4-6

Page 153
Image 153
Siemens SITRANS FUS1010 NEMA-4X/ IP 65 manual Datalogger Control Menu Structure