Siemens SITRANS FUS1010 NEMA-4X/ IP 65 manual Manual Adjustment Procedure

Models: SITRANS FUS1010 NEMA-4X/ IP 65

1 232
Download 232 pages 51.98 Kb
Page 110
Image 110

Section 2

1010NFM-3J

 

 

 

SiemensDual Path 4SS10G

Access Path/Channel Diagnostic Data

Channel /Path Setup

Pipe Data

Application Data

Pick/Install Xdcr

Operation Adjust

Flow/Total Units

Data Span/Set/Cal

Stripchart Setup

Datalogger Setup

I/O Data Control

Diagnostic Data

Dual Path Flow

 

Siemens

Dual Path

4SS10G

 

 

Transducer Setup Data

 

 

 

 

 

 

 

 

 

 

Path Select

1

 

 

 

 

Path Enable

Yes

 

 

 

 

Flow Rate

 

 

 

 

 

Application Info

 

 

 

 

 

Liquid Data

 

 

 

 

 

Site Seup Data

 

 

 

 

 

Test Facilities

 

 

 

 

 

Print Site Setup

No

 

 

 

 

Date Site Created:

04.09.0314.52.23

 

 

 

 

 

 

 

 

 

Diagnostic Data

zIn the [Diagnostic Data] Menu, highlight [Path Select] and select the desired transducer path. Press <ENT> to select path.

zPress the <Down Arrow> and scroll to the [Site Setup Data] menu cell. Press the <Right Arrow> to select it.

Manual” Adjustment Procedure

zIn the [Site Setup Data] Menu, press the <Down Arrow> and scroll to the [HF] menu cell. Press the <Right Arrow> and a pop-up [Manual] prompt will appear as shown below. Note: Press the <Up/Down Arrow> to select [Automatic], if desired.

zUse the numerical keys to input the desired correction value. Press <ENT> to input value.

 

Siemens

 

Dual Path

4SS10G

 

 

 

 

 

 

 

 

 

 

 

 

fx (drive)

29

 

 

 

N (burst length)

5

 

 

 

Ltn in

 

5.316

 

 

 

 

 

 

 

Vf max CU

 

Adjustment

 

 

 

Vs max m/s

 

>Manual

 

 

 

Vs min m/s

 

1355.00

 

 

 

 

 

 

 

Samples/Cycle

29

 

 

 

HF

>-0.000

 

 

 

 

 

 

 

 

Site Setup Data

 

Siemens

 

Dual Path

4SS10G

 

 

 

 

 

 

 

 

 

 

 

 

fx (drive)

29

 

 

 

N (burst length)

5

 

 

 

Ltn in

 

5.316

 

 

 

 

 

 

 

Vf max CU

 

>Manual

 

 

 

Vs max m/s

 

= -0.120

 

 

 

Vs min m/s

 

1355.00

 

 

 

 

 

 

 

Samples/Cycle

29

 

 

 

HF

>-0.000

 

 

 

 

 

 

 

 

Site Setup Data

zThe new correction value will appear next to the [HF] menu cell as shown below.

2-70

Page 110
Image 110
Siemens SITRANS FUS1010 NEMA-4X/ IP 65 manual Manual Adjustment Procedure, Access Path/Channel Diagnostic Data