Section 7

1010NFM-3J

 

 

 

Press the <Down Arrow> from High Limit to advance to the Low Limit cursor control. The <+> and <-> keys are active in this position. Refer to the “Available Adjustments To Spectra Graph” table found in paragraph 7.12.

Press the <Down Arrow> from Low Limit to advance to the Carrier FX control. The <+> and <-> keys are active in this position.

Press the <Down Arrow> from Carrier FX to advance to the Diagnostic Data Display. The <+> and <-> keys are active in this position.

To exit the Spectra Graph, press the <Menu> key.

To display the Digital Flow Display, press the <Menu> key from any Installation Menu location.

7.12AVAILABLE ADJUSTMENTS TO SPECTRA GRAPH

The table below describes the available adjustments that can be made to the Spectra Graph.

ADJUSTMENT

FUNCTION

NOTE

 

 

 

Noise Level Set

Noise Filter

Set to approximately one quarter the Peak

 

 

FFT value for typical flow conditions.

 

 

 

High Limit

Low Pass Filter Setting

Set to two times the maximum flow rate.

 

 

 

Low Limit

High Pass Filter

Set to one half minimum flow if noise is

 

 

present. In most cases, leave at zero.

 

 

 

Carrier Fx

Transmit Frequency

Adjust only under guidance of technical sup-

 

Code

port staff.

 

 

 

7-7