Section 2 |
ZeroMatic (optional function)
NOTE: ZeroMatic is used in the Reflect Mode only. Invoking ZeroMatic clears any existing fixed zero offset.
Use this menu cell to select the ZeroMatic option. If conditions permit the use of the Auto Zero function then the ZeroMatic option can be used as well.
To select and enable the ZeroMatic option:
In the Pick/Install Xdcr menu, press to scroll to the [Zero Flow Adjust] menu cell. To access the [Zero Flow Adjust] option list press
NOTE: If ZeroMatic is not running, the [Actual Zero] menu item will be displayed next to the [Zero Flow Adjust] menu cell.
| Siemens | 2 Channel | SITE1 | |||
| Conforms Indicated flow to Actual Zero |
| ||||
|
|
|
| |||
| Transducer Model | 1011 Universal |
| |||
| Transducer Size |
| B3 |
|
|
|
| Xdcr Mount Mode | Reflect |
|
|
| |
| Spacing Offset |
| Minimum Flow |
| ||
| Number Index |
| 7 |
|
|
|
| Spacing Method |
| Track 1012TP |
| ||
| Ltn Value (in) |
| ReversaMatic |
|
| |
| Install Completed? | ZeroMatic |
|
|
| |
| Empty Pipe Set |
| Auto Zero |
|
|
|
| Zero Flow Adjust |
| Actual Zero |
|
| |
|
|
| ZeroClr |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Pick/Install Xdcr |
|
|
|
|
|
|
|
|
|
|
|
|
Select the [ZeroMatic] menu cell by pressing or then press ENT
| Siemens | 2 Channel | SITE1 | ||||
|
| ZeroMatic Active | [6:.................:0] |
|
| ||
|
|
|
|
| |||
|
| Transducer Model | 1011 Universal |
| |||
|
| Transducer Size | B3 |
|
|
| |
|
| Xdcr Mount Mode | Reflect |
|
|
| |
|
| Spacing Offset |
| Minimum Flow |
| ||
|
| Number Index |
| 7 |
|
|
|
|
| Spacing Method | Track 1012TP |
| |||
|
| Ltn Value (in) |
| 0.778 |
|
|
|
|
| Install Completed? | Yes | ZeroMatic |
| ||
|
| Empty Pipe Set |
| Channel Not Setup |
| ||
|
| Zero Flow Adjust | ZeroMatic |
|
| ||
|
|
|
|
|
|
|
|
Pick/Install Xdcr
When the initial makeup of ZeroMatic is complete the screen will return to the Pick/Install Xdcr menu and automatically highlight [Operation Adjust], which is the next menu cell.