Section 1

1010NFM-3J

 

 

 

1.8USING FASTSTART SETUP

The dual-channel clamp-on mode offers a choice of using FastStart Setup or Full Site Setup to create your site setup. As its name implies, FastStart Setup is a quick way to start up the meter. FastStart inserts default settings into all applicable menu cells. However, these can be edited later via the Full Site Setup menu. In order to use FastStart Setup:

z

z

Your pipe must be included on the meter’s pre-set pipe table.

Your liquid is water, or has similar sonic characteristics.

FastStart setup is actually a shortcut to these two menu locations (where data entry is mandatory for a successful start-up):

z

z

Pipe Data: The Pipe Data Menu is limited to the selection of pipe pre-sets only. There are no provisions for manual pipe editing.

Pick Install Xdcr: This is the same menu as Full Menu Setup (see paragraph 1.7).

1.8.1 CHOOSING THE PIPE CLASS/SIZE

Press to access [FastStart Setup]. This selects [Choose Pipe Class]. There are over sixty English and metric pipe sizes for eight different classes of pipes. Picking a pipe class and size automatically sets all the required pipe parameters using industry standard dimensional data.

Pressto access the Pipe Class option list. Note that although [Manual Entry] exists on this list, it is inactive for FastStart Setup (see paragraph 2.2.1).