Manual Addendum |
|
SETUP PROCEDURE FOR
Caution: DO NOT use the field manual installation procedure to startup a
1.INTRODUCTION
When the system 1010 is
The flow calibration report issued with each
NOTE: In order for the flow calibration to be valid, the flow computer and transducers being used must have the same serial numbers as those listed for the site on the Calibration Diagnostic Data Sheet.
2.SETUP PROCEDURE
2.1Transducer Installation
2.1.1Refer to the diagnostic data sheet to find the mounting mode (Direct or Reflect) used during the
2.1.2Refer to the diagnostic data sheet for the transducer spacing index utilized during the flow calibration. Using the mounting configuration employed during the flow calibration, install the transducers on the pipe at the above noted spacing positions in accordance with the instructions provided on the transducer installation drawings.
2.1.3Attach the transducer cables noting that the cable marked “UP” attaches to the trans- ducer closest to the source of flow.
NOTE: Before proceeding further, ensure that the pipe is full of liquid. It is not impor- tant at this point that it be flowing.
1 | July 2002 |