Section 4 |
| |
|
|
|
|
|
|
When you are sure that the RTD Sensor is at 32ºF (0ºC), press | ENT | to recalibrate the RTD sensor. To |
verify the calibrated reading, go to [Diagnostics/Liquid Data] to check the current RTD output. Make sure that it coincides with the reading of the reference thermometer. Repeat for the other RTD if nec- essary.
Calib @ deg F =32.0
4.7 THE CLOCK SET MENU
The Clock Set menu allows you to set the time and date. The meter uses its internal clock/calendar to record the
CLOCK SET MENU STRUCTURE
Clock Set
Date | xx.xx.xx (date entry) |
Time | xx.xx (time entry) |
Notice:
All flowmeters include a
4.7.1 DATE
The Date command sets the month, day and year for the meter’s internal clock/calendar. Enter the date using the [MM.DD.YY] format. Replace MM with two digits to indicate the month. Replace DD with two digits to indicate the day. Replace YY with two digits to indicate the year. Type the period (.) sepa- rator between each set of digits. For example, you would enter December 7, 1997 by typing 12.07.97. You can use the <Right Arrow> to move the cursor back to a number if you need to
To set the Date:
To enable numeric entry press
Use the Numeric keys to type the date (MM.DD.YY).
To store the date press ENT . This moves the cursor to [Time].
4.7.2 TIME
The Time command sets the hours and minutes for the meter’s internal clock/calendar. Enter the time using the [HH.MM.SS] format. Replace HH with 2 digits to indicate the hour (use