Section 2

 

 

 

1010NFM-3J

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Siemens

2 Channel

SITE 1

 

 

 

Conforms operation to user Preferences

 

 

Channel/Path Setup

Pipe Data

Application Data

Pick/Install Xdcr

Operation Adjust

Flow/Total Units

Data Span/Set/Cal

Stripchart Setup

Datalogger Setup

I/O Data Control

Diagnostic Data

Dual Channel Flow

To disable the ZeroMatic function:

Select the [Pick/Install Xdcr] menu cell from the Dual Channel Flow menu.

Scroll down to the [Zero Flow Adjust] menu cell by pressing

NOTE: The highlighted [ZeroMatic] menu item is the only indication that ZeroMatic is function- ing.

Invoke the ZeroMatic initial makeup procedure as previously described above.

While ZeroMatic initial makeup is running, press

 

to abort the process thereby disabling the

function.

 

 

 

 

The screen will return to the Dual Channel Flow menu and highlight the [Operation Adjust] menu cell.

2.5THE OPERATION ADJUST MENU

This menu becomes available after picking a meter type and measurement channel. We recom- mend that you use it after the transducers are installed and operating to “fine-tune” the meter’s output characteristics.

Each application presents different data display and output requirements due to unique pipe and liquid conditions. Use the Operation Adjust menu to match meter operation to the site. You can set damping controls for the primary flow rate output. You can define a Deadband, (usually a very low flow rate), below which, the flow output will be forced to zero. You can also select the meter response to a continuous Fault condition.

Use this menu cell to select either SmartSlew™ or Time Average output damping.

Use this menu cell to enter a flowrate deadband threshold. Flowrates below this level will be reported as zero flow.

Use this menu cell to change the Reflexor response to a Zero Fault indication by selecting [Zero] in the [Zero/Fault Set] menu.

.

Siemens

2 Channel

[1]

BCD

Determine Fault Memory Delay

 

Damping Control

Smart Slew

Deadband Control

0.000

 

Memory/Fault Set

Memory

Memory Delay <sec>

120

 

Reflexor Zero/Fault Set

Fault

 

Operation Adjust

 

 

 

 

2-34

 

Use this menu cell to select meter response to a fault condition. Fault drops outputs to zero. Memory retains last valid reading for duration of Memory delay.

Use this menu cell to enter the duration (in sec.) that you want the meter to maintain its last reading for during a fault condition. The hidden cell only appears if you select [Memory Mode].

Page 74
Image 74
Siemens SITRANS FUS1010 NEMA-4X/ IP 65 manual Operation Adjust Menu