Section 2 |
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Siemens | 2 Channel | SITE 1 | ||
|
|
| Conforms operation to user Preferences |
|
|
Channel/Path Setup
Pipe Data
Application Data
Pick/Install Xdcr
Operation Adjust
Flow/Total Units
Data Span/Set/Cal
Stripchart Setup
Datalogger Setup
I/O Data Control
Diagnostic Data
Dual Channel Flow
To disable the ZeroMatic function:
Select the [Pick/Install Xdcr] menu cell from the Dual Channel Flow menu.
Scroll down to the [Zero Flow Adjust] menu cell by pressing
NOTE: The highlighted [ZeroMatic] menu item is the only indication that ZeroMatic is function- ing.
Invoke the ZeroMatic initial makeup procedure as previously described above.
While ZeroMatic initial makeup is running, press |
| to abort the process thereby disabling the |
function. |
|
|
|
|
The screen will return to the Dual Channel Flow menu and highlight the [Operation Adjust] menu cell.
2.5THE OPERATION ADJUST MENU
This menu becomes available after picking a meter type and measurement channel. We recom- mend that you use it after the transducers are installed and operating to
Each application presents different data display and output requirements due to unique pipe and liquid conditions. Use the Operation Adjust menu to match meter operation to the site. You can set damping controls for the primary flow rate output. You can define a Deadband, (usually a very low flow rate), below which, the flow output will be forced to zero. You can also select the meter response to a continuous Fault condition.
Use this menu cell to select either SmartSlew™ or Time Average output damping.
Use this menu cell to enter a flowrate deadband threshold. Flowrates below this level will be reported as zero flow.
Use this menu cell to change the Reflexor response to a Zero Fault indication by selecting [Zero] in the [Zero/Fault Set] menu.
.
Siemens | 2 Channel | [1] | BCD |
Determine Fault Memory Delay |
| ||
Damping Control | Smart Slew | ||
Deadband Control | 0.000 |
| |
Memory/Fault Set | Memory | ||
Memory Delay <sec> | 120 |
| |
Reflexor Zero/Fault Set | Fault |
| |
Operation Adjust |
|
| |
|
|
|
Use this menu cell to select meter response to a fault condition. Fault drops outputs to zero. Memory retains last valid reading for duration of Memory delay.
Use this menu cell to enter the duration (in sec.) that you want the meter to maintain its last reading for during a fault condition. The hidden cell only appears if you select [Memory Mode].