Section 2 | ||
|
|
|
To adjust the Tn mark position press the <3> key on the keypad to bring up the [TnSet] command. Pressing the <+> or
Zero Crossover Adjustment: (Hot Key 4)
Observe the small “X” mark located on the zero crossing line near the middle of the receive signal in the Graph Screen above. This “X” indicates the central crossover which the 1010 is using to measure the
Envelope Threshold Adjustment: (Hot Key 5 & 6)
Pressing the <=> key causes the graph to toggle between the default signal waveform screen and the signal envelope screen (see example below). This envelope screen can aid in the diagnosis of Tn errors caused by unusual receive waveform distortion. Signal distortion is sometimes caused by poor transducer selection or poor pipe wall conditions, which may result in an incorrectly measured fluid sound velocity. To improve the automatic measurement of Tn, the envelope threshold limit can be adjusted to exclude portions of the envelope, which may be causing the Tn detection problem.
F | 0.73 |
|
| ||
VS | 1469.73 |
| Envelope Signal | ||
[ ] | 1691:1929 | 476 | |||
| |||||
D | 4:4 4 |
|
| ||
Tn | 55.06 55.06 |
| |||
dT | 3.11 |
|
| ||
S | 72 | 1 91 |
| Envelope Threshold | |
SN | 124:1 3 |
|
| ||
| TN marker |
| Crossover Marker | ||
|
|
| |||
Command Line appears here |
| ||||
| HiSet |
|
|
If it appears that the default placement of the Tn marker is incorrect or unstable, it can be adjusted by pressing the <5> key on the keypad to invoke the [EnvSet] command (while viewing the envelope screen). A horizontal line representing the envelope threshold level will appear along with a number indicating the percentage level. The High and Low thresholds can then be moved either up or down on the envelope using the <+> or