Manual Addendum

1010FMA-4

3.TRANSFER INSTALL FUNCTION

All 1010 flowmeter operating systems (version 3.00.20 and greater) include the installation facil- ity called “Transfer Install.” This function permits the transducers to be repositioned while main- taining all calibration parameters and operation established during the water calibration. The Transfer Install function allows the transducers to be optimally positioned for a different fluid, without the need for a new Initial Makeup procedure.

NOTE: Prior to performing Transfer Install make sure that the water calibration proce- dure was performed and a saved active site exists.

3.1Transfer Install Procedure

To initiate the Transfer Install function, proceed as follows:

3.1.1In the Application Menu press the <Right Arrow> to select the [Liquid Class] menu cell. Scroll and highlight [Estimated Vs m/s] from the option list.

3.1.2Use the numeric keys to change the Estimated Vs to the Estimated Vs value of the customer selected liquid.

3.1.3To enter new Estimated Vs value press <ENT>.

3.1.4Proceed to the Pick/Install Xdcr menu and select the same transducer, mounting mode and spacing offset that was selected for the water calibration.

3.1.5Re-space the transducers to the index position indicated by the flowmeter.

3.1.6Scroll to the [Install Completed?] menu cell and select [Transfer Install] from the option list.

NOTE: If [Transfer Install] does not appear in the option list then either the Estimated Vs or the transducer size was improperly entered. In this case, recall the water calibration site and start the procedure again at Step 3.1.1 above.

3.1.7For MultiPath systems repeat Step 3.1.6 above for the remaining paths.

3.1.8The flowmeter should now be operational at the new spacing location.

NOTE: Depending on the size of the pipeline, the change in the estimated sonic veloc- ity (Vs) and the repositioning of transducers, the flowmeter may not operate out of Fault even if the spool or pipe is filled with liquid. This can be expected when performing a Transfer Install for liquified gases or for clamp-on natural gas flow- meters.

3.2Saving New Transfer Install Site

3.2.1To save the Transfer Installed site, scroll to the Channel Setup menu and press the <Right Arrow>. Press the <Right Arrow> again to select the [Save/Rename] menu cell.

3.2.2Use the numeric keys to rename the Transfer Installed site with the same site name used in Step 3.1.2 above, but with a “T” appended to the end of the site name (e.g., 3CS40T).

3.2.3Press <ENT> to store data.

3

July 2002

Page 8