Section 2 |
| |||
|
|
|
|
|
|
|
|
| |
| To enable numeric entry press | ENT |
| |
|
|
|
|
|
Use the numeric keys to type the required % correction for point #2. Press ENT .
Repeat this procedure for all required calibration points. After entering the last Flow
Calibration Point press to exit MultiPoint calibration.
2.8THE STRIPCHART SETUP MENU
This menu becomes available after picking a Meter Type and Measurement Channel. Use Stripchart Setup to select a data category, the type of scaling and the update interval for the
Use this menu cell to select a data item to display on the stripchart.
Use this menu cell to select whether the stripchart shows data as either a rate or a percentage of span.
Siemens | Dual Path | SITE1 |
| |
| Select Data Units or Percent Display |
| ||
|
|
|
|
|
| Select Data |
| LiquIdent |
|
| Data Display |
| Data Rate Units |
|
| Time Base |
| 1 second |
|
| Stripchart Clear |
| Yes |
|
Use this menu cell to clear the current stripchart.
Stripchart Setup
Use this menu cell to select an update interval for the stripchart.
THE STRIPCHART SETUP MENU STRUCTURE
Stripchart Setup | Select Data | Vfo | ||
|
|
| Vfab | |
|
|
| Visc | |
|
|
| Temp | |
|
|
| Vs | |
|
|
| Valc | |
|
|
| Vaer | |
|
|
| Vsg | |
|
| Data Display | Data Rate Units | |
|
|
| Percent of Span | |
|
| Time Base | 1 Second | |
|
|
| 3 seconds | |
|
|
| 6 seconds | |
|
|
| 12 seconds | |
|
|
| 24 seconds | |
|
|
| 1 minute | |
|
|
| 3 minutes | |
|
|
| 6 minutes | |
|
|
| 12 minutes | |
|
|
| 24 minutes | |
|
|
| 1.2 hours | |
|
| Stripchart Clear | Yes |
|
|
|
| No |
2.8.1 SELECT DATA
The [Select Data] option list allows you to select a data function for the